Repairing the Sampler
4
1100 Series WS MWS Reference Manual
129
Interface Board
1
Switch off the sampler at the main power switch.
2
Disconnect all cables from the existing interface board. Then loosen the
interface board holding screws and slide the board out of its holding rails.
3
Identify the interface board slot cover. Loosen the two holding screws, and
remove the cover.
4
Carefully insert the new interface board into the holding rails, and push the
board into the slot. Make sure the board plugs into the socket correctly.
5
Reconnect all cables to the new interface board.
6
Switch on the sampler.
When required
For all repairs inside the sampler or for installation of the board.
Tools required
Flat-head screwdriver.
Parts required
Interface board, see
C A U T I O N
The interface board is sensitive to electrostatic discharge. Always use the ESD strap
when handling electronic boards.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......