NADC - Specific Error Messages
[300 to 399]
An error number in the range [300 to 399] indicates the instrument has
detected an error relating to the NADC personality.
NADC - Specific Error Message Descriptions
(Number)
Description
(300)
Sync word not found
In an EVM measurement, the sync word is not found
and the synchronization cannot be established when
Sync Word
is selected in the
Burst Sync
menu.
(301)
Valid NADC burst not found
A valid NADC burst is not found when the
Device
is
MS.
(302)
Signal too noisy
The valid EVM measurement cannot be performed,
because the input signal is too noisy.
(303)
Burst Delay exceeds 2 ms limit for EVM
In an EVM measurement, the
Burst Delay
value must be
less than 2 ms.