186
Chapter 6
If You Have a Problem
Error Message Descriptions
Query Error Messages
[
−
499 to
−
400]
An error number in the range [
−
499 to
−
400] indicates the instrument
has found a problem when trying to respond to a SCPI query. The
occurrence of any error in this class will cause the error query bit (bit 2)
to be set in the event status register. If a query error occurs one of the
following is true:
• An attempt is being made to read data from the output queue when
no output is either present or pending.
• Data in the output queue has been lost.
Query Error Message Descriptions
(Number)
Description
(-440)
Query UNTERMINATED after indefinite response
Indicates that a query was received in the same
program message after a query requesting an indefinite
response was executed (see IEEE 488.2, 6.3.7.5).
(-430)
Query DEADLOCKED
Indicates that a condition causing a DEADLOCKED
query error occurred (see IEEE 488.2, 6.3.1.7) (for
example, both the input buffer and the output buffer
are full and the device cannot continue).
(-420)
Query UNTERMINATED
Indicates that a condition causing an
UNTERMINATED query error occurred (see IEEE
488.2, 6.3.2.2) (for example, the device was addressed to
talk and an incomplete program message was received).
(-410)
Query INTERRUPTED
Indicates that a condition causing an INTERRUPTED
query error occurred (see IEEE 488.2, 6.3.2.7) (for
example, a query was followed by DAB or
GET
before a
response was completely sent).
(-400)
Query Error
This is a generic query error for devices that cannot
detect more specific errors. The code indicates only that
a query error as defined in IEEE 488.2, 11.5.1.1.7 and
6.3 has occurred.
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...