Chapter 5
Functional Testing
Residual Responses
4. Configure the test equipment as shown in Figure 5-6, “Residual
5. Perform the following steps for each frequency setting listed in Table
a. Set the center frequency by pressing
FREQUENCY
,
Center Freq
,
“n”, and
MHz
. Where “n” is the frequency value in the Center
Frequency column.
For example:
Press
FREQUENCY
,
Center Freq
, 50, and
MHz
.
b. Set the frequency span by pressing
SPAN
,
Span
, “n”, and
MHz
.
Where “n” is the frequency value in the Span Frequency column.
For example:
Press
SPAN
,
Span
, 5, and
MHz
.
Table 5-20
UUT Initialization
Parameter
Setting
Mode
Basic
Measure
Spectrum
Frequency
Center
50 MHz
Span
5 MHz
Resolution Bandwidth
10 kHz
Attenuation
0 dB
Average
On/Off
On
Mode
Exp
Type
Log-Pwr Avg
Number
25
ADC Rang
Auto/On/Off
Auto
ADC Dither
On/Off
Off
Pre ADC BPF
On/Off
On
Trace Display
Average
Marker 1
Spectrum Avg
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...