Chapter 5
171
Functional Testing
Residual Responses
Procedure
1. If the auto alignment for the UUT has not been run within the last
hour, run it by pressing
System
,
Alignment
, and
Align All Now
2. Preset the UUT by pressing
Preset
.
3. Initialize the parameters for the instrument, as listed in Table 5-20,
“UUT Initialization,” by performing the following steps:
a. Press
MODE
and
Basic
.
b. To expand the Spectrum window to full screen, ensure that the
Spectrum window is highlighted and press
Zoom
.
If the Spectrum window is not highlighted, press
Next Window
until the Spectrum window is highlighted and then press
Zoom
.
c. Press
Marker
,
Trace
, and
Spectrum Avg
.
d. Press
FREQUENCY
,
Center Freq
, 50, and
MHz
.
e. Press
SPAN
,
Span
, 5, and
MHz
.
f. Press
Meas Setup
,
More
until (2 of 2) is displayed,
Advanced
,
More
until (2 of 2) is displayed, and
ADC Dither
until Off is underlined.
g. Press
Meas Setup
,
More
until (2 of 2) is displayed,
Advanced
,
ADC Range
, and
Auto
.
h. Press
View/Trace
,
Trace Display
, and
Average
.
i. Press
Amplitude
,
Ref Value
, 20, and
dB
.
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...