106
Chapter 4
Making Measurements
Making the Adjacent Channel Power (ACP) Measurement
Making the Measurement
NOTE
The factory default settings provide a good starting point. For special
requirements, you many want to change some of the settings. Press
Meas Setup
,
More
,
Restore Meas Defaults
at any time to return all
parameters for the current measurement to their default settings.
Select the desired center frequency.
Press
Measure
,
ACP
to immediately make Adjacent Channel Power
(ACP) the active measurement.
To change any of the measurement parameters from the factory default
values, refer to the “Changing the Measurement Setup” section below.
Results
The following figure shows an example result of
ACP (Total Pwr Ref)
measurements in the bar graph window. The absolute and relative
power levels on both sides of the carrier frequency are displayed in the
graphic window and text window.
Figure 2
Adjacent Channel Power Measurement - Bar Graph View