Chapter 5
167
Functional Testing
Phase Noise
5. Configure the test equipment as shown in Figure 5-5, “Phase Noise
6. Perform the following steps to establish the source reference
amplitude:
a. Set the signal source frequency to 600 MHz.
b. If the trace is not on the display, adjust the amplitude reference
value by pressing
Amplitude
and
Ref Value
. Then use the RGB
knob to adjust the reference value to bring the trace onto the
display.
c. Press
Search
.
d. Adjust the source amplitude until the UUT measures
7 dBm
±
0.5.
e. Press
Restart
.
f. Wait for the transmitter tester to finish averaging.
g. Read the marker value in dBm.
h. Record the measured value as Y
i ref
.
Y
i ref
= ___________ dBm
Table 5-16
Synthesizer Signal Generator Initialization
Parameter
Setting
Frequency
600 MHz
Function
Sine Wave
Amplitude
7 dBm