Chapter 4
97
Making Measurements
Making the Spectrum (Frequency Domain) Measurement
Two marker lines are activated at the extreme left side of the
horizontal scale. Press
Normal
and move marker 1 to the desired
place by rotating the
RPG
knob.
Press
Delta
to bring marker 2 to the same place as marker 1.
Move marker 1 to the other desired position by rotating the
RPG
knob. Band power measures the average power between the two
markers. When the band power markers are active, the results are
shown in the results window as Mean Pwr (Between Mks). When the
band power function is off the results window reads Mean Pwr
(Entire Trace).
Troubleshooting Hints
Changes made by the user to advanced spectrum settings, particularly
to ADC range settings, can inadvertently result in spectrum
measurements that are invalid and cause error messages to appear.
Care needs to be taken when using advanced features.