130
Chapter 4
Making Measurements
Timebase Frequency
Timebase Frequency
Purpose
In the absence of a frequency counter, the measurement quickly
determines the frequency difference between the instrument’s 10 MHz
reference oscillator signal and an accurate, external 10 MHz frequency
standard.
NOTE
The timebase frequency check is not as accurate as a measurement of
the
10 MHz Out
signal (rear panel) using a frequency counter
referenced to a frequency standard.
Measurement Method
The internal 10 MHz reference signal is compared with a reliable
external 10 MHz signal which the user inputs into the front panel RF
connector. The transmitter tester subtracts its 10 MHz reference from
the input signal and displays the frequency difference.
Test Setup
An accurate external 10 MHz frequency signal, such as a house
standard, is connected to an attenuator, which attenuates the signal to
0 dB. The attenuated 10 MHz signal is connected to the RF input of the
transmitter tester. Pressing
Mode
,
Service
,
Measure
, Timebase
Frequency will result in the display of data described below.
NOTE
When setting up this test ensure that the reference is set to internal. To
check this press
System
,
Reference
. The
Freq Ref
softkey allows you to
toggle between
Int
(internal) and
Ext
(external). Make sure
Int
is
underlined.
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...