142
Chapter 5
Functional Testing
Getting Started
Table 5-1
Required Test Equipment Summary
Instrument
Critical Specifications
Recommended
HP/Agilent
Model Number
Alternative
HP/Agilent
Model Number
Signal Source
Synthesized
Signal Generator
Frequency:
7 MHz to 2.5 GHz
Spectral purity: 5 to
1321 MHz
SSB Phase Noise:
≤−
110 dBc/Hz at 10 kHz
offset and
≤−
120 dBc/Hz at
30 kHz offset.
8663A
8662A
8664A
8665A
8643A
8644A
w/ Option 002
Synthesized
Swept Signal
Generator
Frequency: 10 MHz to 4.0 GHz
Harmonic level: <
−
40 dBc
Amplitude range:
+
15 to
−
20 dBm
83620B
8360 series
Meters
Power Meter
Must be compatible with the
power sensor.
EPM- 441A
(E4418A)
EMP-442A
(E4419A)
437A
438A
Power Sensor
Frequency Range:
7 MHz to 4 GHz.
Power Range:
−
25 dBm to 10 dBm
SWR:
≤
1.3
Type N (m)
8482A
8481A
E4412A
8482A
w/Option H84
Terminations
Termination 50
Ω
Type N (m) Connector
Frequency: 7 MHz to 4 GHz
909A option 012
Miscellaneous Devices
Power Splitter
Frequency:
1 MHz to 4.3 GHz
11667A
11667B
Cables
Type N
2 required
Precision Type N
62 cm (24in) (m)
11500C
8120-4781
Adapters
Type N (f) to N (f)
50
Ω
1250-1472
1250-0777
3.5 mm (f) to N (f)
50
Ω
1250-1745
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...