210
Chapter 6
If You Have a Problem
Error Message Descriptions
W-CDMA - Specific Error Messages
[500 to 599]
An error number in the range [500 to 599] indicates the instrument has
detected an error relating to the W-CDMA personality.
W-CDMA - Specific Error Message Descriptions
(Number)
Description
(501)
Signal too noisy
(502)
Input power too low
(503)
Cannot correlate to input signal
Cannot correlate to the input signal and no active
channel is found. (from composite EVM measurement)
cdma2000 - Specific Error Messages
[600 to 699]
An error number in the range [600 to 699] indicates the instrument has
detected an error relating to the cdma2000 personality.
cdma2000 - Specific Error Message Descriptions
(Number)
Description
(601)
Signal too noisy
(602)
Input power too low
(603)
Can not get long code phase (RS-232)
For MS (mobile station) measurements, the long code
phase information could not be obtained from the signal
at the RS-232 port.(from code domain power
measurement or composite EVM measurement)
(604)
Cannot correlate to input signal
Cannot correlate to the input signal and no active
channel is found. (from composite EVM measurement)
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...