152
Chapter 5
Functional Testing
Amplitude Accuracy at 50 MHz
e. Record the signal amplitude, as measured by the power meter, in
the Amplitude Input
i
column of Table 5-8, “Amplitude Accuracy
f. Press
Restart
.
g. Wait for the transmitter tester to finish averaging, and read the
marker value.
h. Record the signal amplitude, as measured by the UUT, in the
Measured Amplitude X
i
column of
i. Calculate the amplitude accuracy using the following equation
and record in the Error Amp_Acc
i
column of
Error Amp_Acc
i
= Input
i
− Χ
i
Table 5-7
Measurement Settings
UUT Nominal
Measured Values
(dBm)
Nominal Source
Amplitude
(dBm)
UUT Input
Attenuator
(dB)
−
2
5
0
−
12
−
5
0
−
22
−
15
0
2
9
4
−
12
−
5
4
−
22
−
15
4
3
10
5
−
7
0
5
−
17
−
10
5
8
15
10
−
12
−
5
10
−
2
5
21