Chapter 6
If You Have a Problem
Error Message Descriptions
(-321)
Out of memory
Indicates that an internal operation needed more
memory than was available.
If this occurs during a memory catalog display, it means
the system did not have enough free RAM to prepare
the catalog.
(-320)
Storage fault
Indicates that the firmware detected a fault when using
data storage. This error is not an indication of physical
damage or failure of any mass storage element.
(-315)
Configuration memory lost
Indicates that non-volatile configuration data saved by
the device has been lost. The meaning of this error is
device-dependent.
(-314)
Save/recall memory loss
Indicates that the non-volatile data saved by the
*SAV?
command has been lost.
(-313)
Calibration memory lost
Indicates that non-volatile calibration data has been
lost.
(-312)
PUD memory lost
Indicates that the protected user data saved by the
*PUD
command has been lost.
(-311)
Memory error
Indicates that an error was detected in the device’s
memory.
(-310)
System error
Indicates that an error, termed “system error” by the
device, has occurred.
(-300)
Device-specific error
This is a generic device-dependent error for devices
that cannot detect more specific errors. The code
indicates only that a device-dependent error as defined
in IEEE 488.2, 11.5.1.1.6 has occurred.
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...