156
Chapter 5
Functional Testing
Input Attenuator Accuracy at 50 MHz
4. Measure the reference value by performing the following steps:
a. Press
Search
.
b. Record the marker value as
Χ
ref
in the Measured Value X
i
column
of Table 5-10, “Measurement Settings.”
Table 5-9
UUT Initialization
Parameter
Setting
Mode
Basic
Measure
Spectrum
Reference Level
−
23 dBm
Frequency
Center
50 MHz
Span
20 kHz
Resolution Bandwidth
400 Hz
Input Port
50 MHz Ref
Input Attenuation
10 dB
Average
On/Off
On
Mode
Exp
Type
Log-Pwr Avg
Number
25
Trace Display
Average
Marker
Spectrum Avg