Chapter 5
173
Functional Testing
Residual Responses
c. Set the resolution bandwidth by pressing
Meas Setup
,
Res BW
,
“n”, and
kHz
. Where “n” is the frequency value in the Res BW
column.
For example:
Press
Meas Setup
,
Res BW
, 10, and
kHz
.
d. Press
Search
.
e. Wait for the instrument to finish averaging.
f. Read the marker value in dBm.
g. Record the measured value as
Χ
i
in the Measured X
i
column of
Table 5-22, “Residual Response Report.”
Table 5-21
Measurement Settings
Center Frequency
(MHz)
Span Frequency
(MHz)
Res BW
(kHz)
50.00
5
10
740.00
2
7
1804.00
2
7
1917.90
2
7
1990.80
2
7
1992.90
2
7
3768.30
2
5
3899.50
2
5
3900.00
2
5
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...