Chapter 5
151
Functional Testing
Amplitude Accuracy at 50 MHz
4. Preset the signal source and initialize the test equipment
parameters as listed in Table 5-6, “Signal Source Initialization.”
5. Configure the equipment as shown in Figure 5-3, “Amplitude
6. Perform the following steps for each of the nominal measured values
listed in Table 5-7, “Measurement Settings”:
a. Set the signal source amplitude to the value listed in the Nominal
Source Amplitude column in
b. On the UUT, press
Restart
and
Search.
c. Wait for the transmitter tester to finish averaging.
d. Adjust the source amplitude, as measured on the UUT, to
±0.5
dBm of the value listed in the Nominal Measured Value
column in
Table 5-5
UUT Initialization
Parameter
Setting
Mode
Basic
Measure
Spectrum
Frequency
Center
50 MHz
Span
1 MHz
Resolution Bandwidth
20 kHz
Input Attenuation
0 dB
Average
On/Off
On
Mode
Exp
Type
Log-Pwr Avg
Number
10
Trace Display
Average
Marker
Spectrum Avg
Table 5-6
Signal Source Initialization
Parameter
Setting
Frequency
50 MHz
Level
5
dBm
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...