166
Chapter 5
Functional Testing
Phase Noise
4. Preset the signal source and initialize the parameters as shown in
Table 5-16, “Synthesizer Signal Generator Initialization.”
Table 5-15
UUT Initialization
Parameter
Setting
Mode
Service
Measure
Spectrum
Frequency
Center
600 MHz
Span
800 Hz
Resolution Bandwidth
Auto
Attenuation
10 dB
Average
On/Off
On
Mode
Exp
Type
RMS
Number
5
ADC Range
Manual
−
6
ADC Dither
On/Off
On
Pre ADC BPF
On/Off
Off
Trace Display
Average
Marker 1
Spectrum Avg