108
Chapter 4
Making Measurements
Making the Adjacent Channel Power (ACP) Measurement
Changing the Measurement Setup
Table 4-1
Adjacent Channel Power Measurement Defaults
Measurement Parameter
Factory Default Condition
View/Trace
Bar Graph (Total Pwr Ref)
Spectrum Trace
On
Average Number
20 On
Average Mode
Repeat
Reference Channel
Channel Integ BW
Avg Type
1.23000 MHz
Pwr Avg (RMS)
Offset Limits
Offset
Offset Freq
Offset Side
Ref BW
Avg Type
A
750.000 kHz On (offset A)
Both
30.000 kHz
Pwr Avg (RMS)
Limit Setup
Abs Limit
Fail
Rel Lim (Car)
Rel Lim (PSD)
0.00 dBm
Relative
-45.00 dBc (offset A)
-28.87 dB (offset A)
Measurement Type
Total Pwr Ref
Trigger Source
Free Run (Immediate)
Limit Test
On
Chan Integ BW
1.23000 MHz
Meas Type
Total Pwr Ref
Offset & Limits:
Offset
Offset frequency
Reference BW
Absolute limit
Fail
Relative limit (Car)
Relative limit (PSD)
A
750.000 kHz On (offset A)
30.000 kHz
0.00 dBm
Relative
−
45.00 dBc (offset A)
−
28.87 dB (offset A)
Advanced
Ref Chan Advanced
Содержание E4406A
Страница 4: ...4 ...
Страница 10: ...10 Contents ...
Страница 138: ...138 Chapter4 Making Measurements Front Panel Test ...
Страница 139: ...139 5 Functional Testing ...
Страница 214: ...214 Chapter6 If You Have a Problem Returning Your Instrument to Agilent Technologies ...
Страница 215: ...Chapter 6 215 If You Have a Problem Repair Tag Goes Here Remove This Page Repair Tag Goes Here Remove This Page ...
Страница 216: ...216 Chapter6 If You Have a Problem Repair Tag Goes Here Remove This Page ...
Страница 217: ...217 7 Options and Accessories ...
Страница 226: ...226 Chapter7 Options and Accessories Accessories ...