424
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_pferror.fm
SETup:PFERror
SETup:PFERror:TIMeout:TIME
SETup:PFERror:TIMeout:STATe
SETup:PFERror:TRIGger:DELay
Function
Sets/queries the timeout value in seconds that will be used for phase/frequency measurements
when the timeout state is ON. The units (S|MS) are optional, if no units are specified then unit
default to S.
Setting
Range: .1 to 999
Resolution: .1
Query
Range: .1 to 999
Resolution: .1
*RST Setting
10 seconds
Programming Example
OUTPUT 714;"SETUP:PFERROR:TIMEOUT:TIME 4" !Sets the timeout value to 4 seconds.
Function
Selects/queries the timeout state for a phase/frequency measurement.
Setting
0|OFF | 1|ON
Query
0|1
*RST Setting
0|off
Programming Example
OUTPUT 714;"SETUP:PFERROR:TIMEOUT:STATE ON" !Sets the timeout state to on for a
!Phase/Frequency measurement.
Function
Sets/queries the trigger delay time in seconds for a phase/frequency measurement. The units
(S|MS|US|NS) are optional, if no units are specified then units default to S.
See
“Phase and Frequency Error Measurement Description” on page 82
Setting
Range:
−
2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 ns, whichever is greater
Query
Range:
−
2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 ns, whichever is greater
*RST Setting
0 seconds
Programming Example
OUTPUT 714;"SETUP:PFERROR:TRIGGER:DELAY 1.2MS" !Sets trigger delay time to
!1.2 milli-seconds