357
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_initiate.fm
INITiate
INITiate:DONE?
INITiate:DONE:CLEar
INITiate:DONE:FLAG[:ALL]
Function
Queries (one at a time) which measurements if any are available or have timed out.
See
“Measurement Event Synchronization” on page 132
for how to use this command.
See
“Measurement States” on page 150
to understand the test set’s measurement states.
See
“INITiate:DONE:FLAG<measurement mnemonic>” on page 358
for include or exclude
commands
Query
Range: NONE| TXP | PVT | PFER | ORFS | AAUD | DAUD | DPOW | FBER | BERR | IQT
| WAIT
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Clears the done flag from all measurements. See
Programming Example
OUTPUT 714;”INITIATE:DONE:CLEAR” !clears done flag.
Function
Specifies that all measurements are considered, (included or excluded) when the DONE? query is
sent.
If a measurements trigger arm, see
“Trigger Arm (Single or Continuous) Description” on page
, has been left in continuous mode, the done flag for that measurement will toggle between
DONE and WAIT, see
. The INITiate:DONE? query will probably
not be able to catch the measurement at the instant it is done, therefore the measurement will
never appear to be done. If a measurement trigger arm must be left in continuous mode the user
should (exclude) it, using this command, from the INITiate:DONE? query results.
Once the INITiate:DONE:FLAG has been set to EXCLude for a measurement, the user must
send the INCLude command for that measurement in oder to query that measurement with, the
INITiate:DONE? query. The test set will not reset any (excluded measurement) to be an (included
measurement) with any form of preset, see
“Preset Descriptions” on page 535
Setting
Range
• INCLude: include all measurements
• EXCLude: exclude all measurements
Programming Example
OUTPUT 714;”INITIATE:DONE:FLAG:ALL EXCLUDE” !Excludes all measurements from
!contributing the INITIATE:DONE?
!query, see “INITiate:DONE?” on
!page 6.