493
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_SYSTem_PRESet.fm
SYSTem:PRESet
SYSTem:PRESet
“Diagram Conventions” on page 213
SYSTem:PRESet[1] (not recommended for use)
SYSTem:PRESet2
Function
Not recommended for use at this time, use the SYSTEM:PRESET3 command for partial
preset.
Performs a partial preset. This is the recommended command when a user wants to change
from remote operation to manual operation and a partial preset is needed.
Any call in process is disconnected and all measurements are aborted and inactivated.
Measurement parameters are not changed.
A partial preset will not modify any measurement settings including trigger arm. See
“Trigger Arm (Single or Continuous) Description” on page 151
Related Topics
for more details
Programming Example
OUTPUT 714;”SYSTEM:PRESET” !Partial preset when changing from remote to manual
operation.
Function
Performs a full preset the same as using the *RST command. Either command may be used when
a full preset is needed during remote operation of the test set.
All parameters are set to their default values. All measurements are aborted the trigger arm is
set to single. See
“Trigger Arm (Single or Continuous) Description” on page 151
Related Topics
See
Programming Example
OUTPUT 714;”SYSTEM:PRESET2” !Full preset.
SYSTem
[1]