![Agilent Technologies 8960 Series 10 Скачать руководство пользователя страница 411](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627411.webp)
411
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_iq.fm
SETup:IQTuning
SETup:IQTuning:TRIGger:DELay
SETup:IQTuning:TRIGger:SOURce
Function
Sets/queries the trigger delay time in seconds for an I/Q Tuning measurement. The units
(S|MS|US|NS) are optional, if no units are specified then the default is S.
Setting
Range: -2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 ns, whichever is greater
Query
Range: -2.31 ms to +2.31 ms
Resolution: 5 significant digits or 100 ns, whichever is greater
*RST
zero seconds
Programming Example
OUTPUT 714;"SETUP:IQTUNING:TRIGGER:DELAY 1.2MS" !Sets the trigger delay
time to 1.2 milli seconds.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Function
Sets/queries the trigger source for an I/Q Tuning measurement.
Setting
RISE|IMMediate
See
“Triggering of Measurements” on page 149
Query
RISE|IMM
*RST
RISE
Programming Example
OUTPUT 714;"SETUP:IQTUNING:TRIGGER:SOURCE IMM" !Sets the trigger source to
immediate.
XXXXXXXXX
XXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXXX
Содержание 8960 Series 10
Страница 26: ...26 Contents ...
Страница 363: ...363 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual Chapters hpib_read fm READ ...
Страница 500: ...500 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual ch_gen_usage fm General Usage ...