![Agilent Technologies 8960 Series 10 Скачать руководство пользователя страница 215](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627215.webp)
215
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_abort_description.fm
ABORt Subsystem
ABORt Subsystem
Description
The ABORt command causes a measurement cycle in progress to stop. If the measurement is not being
continuously armed (single trigger) , the measurement will remain in the idle state after this event. If the
measurement is being continuously armed, a new measurement cycle will begin after ABORt. If an ABORt
command is issued from any measurement state other than measuring, the command is ignored.
Other Commands that Execute an ABORt Action
INITiate:<meas> will execute an ABORt:<meas> as part of the INITiate:<meas> command.
READ:<meas>? will execute an ABORt:<meas> action that aborts just one trigger sequence and then
combines the INITiate and FETCh? commands.
Syntax Diagram and Command Descriptions
Содержание 8960 Series 10