CALibration
CALibration
April 30, 1999
“Diagram Conventions” on page 213
CALibration:IQ[:1]?
NOTE
When the the calibration is done the test set display will display:
IQ Calibration completed successfully for modulator 1. Cycle power to continue.
Function
Sets/queries the calibration of the IQ modulator for RF generator 1. It takes some time to
complete calibration and can’t be aborted except by cycling the power switch.
• Calibrates the IQ modulator for RF generator 1.
• Returns a value indicating success or failure of calibration.
Query
Range
• 0 = Pass
•
−
340 = Fail
Programming Example
OUTPUT 714;"CALIBRATION:IQ?" !Performs a calibration of the IQ modulator for
!RF generator 1 and returns a value indicating
!success or failure.
CALibration
(returns 0 for pass or -340 for fail)
<sp><YYYY,MM,DD>
?
[:1]
? (returns 0 for pass or -340 for fail)