![Agilent Technologies 8960 Series 10 Скачать руководство пользователя страница 472](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627472.webp)
472
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_STATUS_STD_EVNT.fm
Standard Event Status Register
Program Example - Standard Event Status Register
OUTPUT 714;”*ESR?” !Queries (reads) the Standard Event Status Register.
2
4
Query Error
This bit is set to 1 if an error has occurred while
trying to read the test set’s output queue. The
following events cause a query error:
• An attempt is made to read data from the output
queue when no data is present or is pending.
• Data in the output queue has been lost. An
example of this would be an output queue overflow.
1
2
Reserved for future use.
This bit will always be 0.
0
1
Operation Complete
This bit is set to 1 when the test set has completed all
pending operations and is ready to accept new
commands. This bit is only generated in response to
the *OPC IEEE 488.2 common command.
Bit
Number
Binary
Weighting
Condition
Description