S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_power.fm
CALL:POWer
CALL:POWer
“Diagram Conventions” on page 213
CALL[:CELL]:POWer[:SAMPlitude]
Function
Sets/queries the value for Cell Power and turns the state to ON. This is the same for BCH
and TCH.
The suffix dBm is optional.
The Cell Power field is affected when there is an amplitude offset, see
Setting
range: -10 dBm to -127 dBm
resolution: .01 dBm
Query
range: -10 dBm to -127 dBm
resolution: .01 dBm
*RST Setting
-85 dBm
Programming Example
OUTPUT 714;”CALL:CELL:POWER:SAMPLITUDE -50dBm” !Sets the value to -50dBm
!and the state to ON.
CALL
[:CELL[1]]
<sp>1|ON|0|OFF
? (returns 1|0)
<sp><num value>[dBm]
[:SAMPlitude]
<sp><num value>
?
?