535
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\gen_inst_states.fm
Preset Descriptions
Preset Descriptions
Description
The test set is capable of accepting several different preset commands.
At no time during a preset operation, will transmit power exceed the last user setting of the transmit power.
The input power will not be set to any value lower than the last user setting of the input power. This is to avoid
power spikes on the output and possible receiver damage on the input during transitions associated with
preset operations.
Examine the results in Tables 4 and 5 to determine which preset to use for your situation.
Partial Preset
Users save setup time with a partial preset because measurement setup parameters remain unchanged. This
is the recommended way to place the test set in a known condition.
Press the green PRESET key on the front panel to perform a partial preset.
Example
OUTPUT 714;”SYSTEM:PRESET3” !Command for a partial preset when user in
!remote operation.
SYSTEM:PRESET3 is the recommended command for a Partial Preset operation. The SYSTEM:PRESET[1]
Table 1. Partial Preset Behavior
Function
Result
Trigger Arm
no change
Measurement parameters
no change
Call in progress
aborted
Operating Mode
Active Cell
Measurements
aborted and inactivated
Measurement results
NAN
Cell Activated State
ON
Cell Power State
ON
Call Control Status
Idle
Call Counters
cleared
Call Error Counters
cleared
SACCH
cleared
Measurement integrity
indicator
1 = no result available