Test Mode Operating Mode
Example 11. BCH + TCH Test Function Using Manual Receiver Control
The following example shows how to set up a test mode measurement using the test function BCH + TCH. In
this example the test set transmits a BCH on PGSM channel 21 and a TCH on PGSM channel 31. Manual
receiver control is used to configure the test set to measure a signal from the mobile station at 895 MHz, the
frequency can also be tuned by channel number with the manual channel parameter. Expected power is set at
+14 dBm.
1. Select test mode.
OUTPUT 714;”CALL:OPERATING:MODE TEST”
2. Select PGSM as the broadcast band (the TCH will automatically be set to this band).
OUTPUT 714;”CALL:CELL:BAND PGSM”
3. Select BCH + TCH as the test function.
OUTPUT 714;”CALL:FUNCTION:DOWNLINK BCHTCH”
4. Set the BCH to channel 21.
OUTPUT 714;”CALL:BCH:PGSM 21”
5. Set the TCH to channel 31.
OUTPUT 714;”CALL:TCH:PGSM 31”
6. Configure the receiver control to manual.
OUTPUT 714;”RFANALYZER:CONTROL:AUTO OFF”
7. Configure the test set’s receiver frequency to 895 MHz.
OUTPUT 714;”RFANALYZER:MANUAL:FREQUENCY 895 MHZ”
If tuning by channel number, see
“RFANalyzer:MANual:CHANnel[:SELected]” on page 375
8. Set the receivers expected power level to 14 dBm.
OUTPUT 714;”RFANALYZER:EXPECTED:POWER 14 DBM”
9. To make the measurement, set up the mobile station to transmit at 895 MHz and at a power level of 14
dBm.