![Agilent Technologies 8960 Series 10 Скачать руководство пользователя страница 227](http://html1.mh-extra.com/html/agilent-technologies/8960-series-10/8960-series-10_reference-manual_2868627227.webp)
227
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_call_activated.fm
CALL:ACTivated
CALL:ACTivated
April 20, 1999
“Diagram Conventions” on page 213
CALL[:CELL]:ACTivated[:STATe]
Function
Turns BS Emulator control of all signalling operations, uplink demodulation and downlink (BCH
& TCH) generation on or off. Query form returns a 1 (state = on) or a 0 (state = off).
When the test set’s operating mode is “test mode” or when the cell activated state is “off”, the
burst type may need to be specified before the test set can synchronize to the input signal’s
midamble. See
Setting
range: 1 | ON | 0 | OFF
Query
range: 1 | 0
*RST Setting
1 (state = ON)
Related Topics
“Configuring the Broadcast Channel (BCH)” on page 511
Programming Example
OUTPUT 714;”CALL:CELL:ACTIVATED:STATE OFF” !Turns all signalling operations,
!uplink demodulation and downlink
!(BCH & TCH) generation off.
:ACTivated
[:STATe]
<sp>1|ON|0|OFF
? (returns 1|0)
CALL