393
S:\Hp8960\E1960A GSM Mobile Test Application\A.04 Release\Reference_Manual\Chapters\hpib_setup_fberror.fm
SETup:FBERror
SETup:FBERror:CLSDelay:STATe
SETup:FBERror:CONTinous
Function
Selects/queries the closed loop signalling delay state for Fast Bit Error measurements. If the
state is off the test set will not wait to start a FBER measurement after a downlink signalling
operation has completed.
The delay time defines how long the test set should wait before starting and FBER measurement
after a downlink signalling operation has completed and after the test set has sent a close loop
command to the MS.
When a close loop message is set to the MS the closed loop signalling delay time will hold off the
FBER measurement from starting for the specified time period.
Setting
Range: 1 | ON | 0 | OFF
Query
Range: 1 | 0
*RST
1 | ON
Programming Example
OUTPUT 714;”SETUP:FBERROR:CLSDELAY:STATE ON” ! Sets the Close Loop Delay
! state to on.
Function
Selects/queries the trigger state for Fast Bit Error Rate tests.
Setting
Range:
1 | ON = Continuous trigger mode
0 | OFF = Single trigger mode
Query
Range: 1 | 0
*RST
0 | single
Programming Example
OUTPUT 714;”SETUP:FBERROR:CONTINUOUS 0” !Specifies single trigger mode for Fast
!BER measurements.
Содержание 8960 Series 10
Страница 26: ...26 Contents ...
Страница 363: ...363 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual Chapters hpib_read fm READ ...
Страница 500: ...500 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual ch_gen_usage fm General Usage ...
Страница 574: ...574 S Hp8960 E1960A GSM Mobile Test Application A 04 Release Reference_Manual ch_error_messages fm Error Messages ...