6
Maintaining the Purged Packed Inlet
To Clean the Septum Seat in the Purged Packed Inlet
84
Maintaining Your GC
To Clean the Septum Seat in the Purged Packed
Inlet
1
Gather the following:
•
Replacement septum, see
“Consumables and Parts for the Purged Packed Inlet”
•
Septum nut wrench
•
0- or 00-grade steel wool (optional)
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
•
Ultrasonic cleaning bath
•
Lint-free gloves
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands
.
3
Remove the septum retainer nut or Merlin cap.
4
Loosen the top insert assembly and remove.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
5
Use tweezers to remove the septum or Merlin Microseal from the top insert assembly. Do
not gouge or scratch the interior of the septum head.
6
Scrub the residue from the top insert assembly and septum nut with a small piece of
rolled-up steel wool and tweezers. Ultrasonically clean the retainer nut and top insert
assembly.
7
Use compressed air or nitrogen to blow away the pieces of steel wool and septum.
8
Wearing gloves, inspect the O-ring and replace, if necessary. See
9
Install the top insert assembly and hand-tighten firmly.
10
Firmly press the new septum or Merlin Microseal into the fitting.
11
Install the septum retainer nut or Merlin cap and finger-tighten. Tighten the septum retainer
nut until the C-ring is about 1 mm above the nut.
CAUTION
Overtightening the septum nut can cause contamination.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...