Maintaining the Split/Splitless Inlet
To Replace the Filter in the Split Vent Trap for the Split/Splitless Inlet
Maintaining Your GC
67
To Replace the Filter in the Split Vent Trap for the
Split/Splitless Inlet
1
Gather the following:
•
New split vent filter cartridge. See
“Consumables and Parts for the Split/Splitless
2
Launch the GC maintenance wizard:
Maintenance > Inlets,
select the inlet, then select
Perform Maintenance > Replace Split Vent Trap > Start Maintenance.
The wizard will walk
through the steps needed to replace the filter (those same steps are repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
WARNING
The split vent trap may contain residual amounts of any samples or other chemicals you
have injected into the GC. Follow your company’s safety procedures for handling these types
of substances while replacing the trap filter cartridge.
3
Remove the pneumatics cover (top, back of GC). See
4
Completely loosen the knurled nut that secures the split vent trap in place as shown in
.
Figure 22.
Loosen the knurled nut
5
Slide the trap assembly back from the guided mounting bracket and tilt up to expose the
filter as shown in
.
Knurled nut secured
Knurled nut loosened
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...