Maintaining Capillary Columns
To Reverse a Column and Bakeout Contaminants
Maintaining Your GC
37
To Reverse a Column and Bakeout Contaminants
1
Gather the following:
•
1/4-inch wrench
•
Column cutter
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance
and wait for the GC to become ready
.
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3
Disconnect the column from the inlet and detector.
4
If necessary, cut a loop from the column. See
Do not attach the column to the inlet.
5
Remove the column from the hanger and reverse its position (inlet and detector ends) and
place the column back on the hanger.
6
Attach the column to the inlet.
•
“To Install a Capillary Column with the Split/Splitless Inlet”
•
“To Install a Capillary Column with the Multimode Inlet”
•
“To Install a Capillary Column with the Purged Packed Inlet”
•
“To Install a Capillary Column with the Cool On-Column Inlet”
•
“To Install a Capillary Column with the Programmable Temperature Vaporization
Inlet”
•
“To Install a Capillary Column with the Volatiles Interface”
7
Attach your column to the detector.
•
To Install a Capillary Column in the FID
•
To Install a Capillary Column in the NPD
•
To Install a Capillary Column in the TCD
•
To Install a Capillary Column in the ECD
•
To Install a Capillary Column to the FPD+
8
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
9
Set the column flow to the normal operating value, or set the capillary column gas velocity
to 30 cm/s.
For Multimode, Split/Splitless, PTV, and VI inlets, select split mode and set the split vent
flow to 200 mL/min.
10
Purge the column with carrier flow for at least 10 minutes before heating the oven.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...