Maintaining the PTV Inlet
To Install a Capillary Column with the Programmable Temperature Vaporization Inlet
Maintaining Your GC
147
To Install a Capillary Column with the
Programmable Temperature Vaporization Inlet
1
Gather the following, see
“Consumables and Parts for the Programmable Temperature
•
Column
•
Graphpak 2M ferrule
•
Column nut
•
Column cutter
•
Septum
•
Isopropanol
•
Lab tissue
•
Metric ruler
•
5-mm and 6-mm wrenches
•
Typewriter correction fluid or a marker
•
Lint-free gloves
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3
Before installing the column, install the correct column adapter. See
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
Place the column on the hanger with the ends pointing up and the label to the front.
5
Place a Graphpak 2M ferrule onto the column, with the graphite facing up towards the inlet.
6
Score the column using a glass scribing tool. The score must be square to ensure a clean
break.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...