![Agilent Technologies 8890 Скачать руководство пользователя страница 107](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678107.webp)
7
Maintaining the COC Inlet
To Install a Capillary Column with the Cool On-Column Inlet
Maintaining Your GC
107
6
Break off the column end by supporting it against the column cutter opposite the scribe.
Inspect the end with a magnifying loupe to make certain there are no burrs or jagged
edges.
7
Wipe the column walls with a tissue dampened with isopropanol to remove fingerprints
and dust.
8
Gently insert the column into the inlet until it bottoms. You should feel spring tension as
you push up on the column. (Do not withdraw the column.)
9
Insert the column nut into the inlet fitting and finger-tighten.
WARNING
To avoid bending the inlet, always use two wrenches. Use a 5/16-inch wrench to support the
inlet while tightening the column nut with a 1/4-inch wrench.
10
Tighten the column nut an additional 1/4 turn with a wrench or until the column does not
move.
11
If using an automatic injection system with a 0.25-mm or 0.32-mm column, verify that the
column installation by manually pushing the syringe into the inlet.
Good
Bad
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...