![Agilent Technologies 8890 Скачать руководство пользователя страница 65](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678065.webp)
5
Maintaining the Split/Splitless Inlet
To Replace the Gold Seal on the Split/Splitless Inlet
Maintaining Your GC
65
To Replace the Gold Seal on the Split/Splitless Inlet
1
Gather the following:
•
Replacement gold seal, see
“Consumables and Parts for the Split/Splitless Inlet”
•
Replacement washer
•
1/4-inch wrench (for column)
•
1/2-inch wrench
•
Lint-free gloves
2
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
3
Remove the inlet liner.
4
Remove the column from the inlet. Cap the open end of the column to prevent
contamination. Remove the insulation cup around the base of the inlet.
5
Loosen and remove the reducing nut. Remove the washer and seal inside the reducing nut.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
Remove insulation cup
Disconnected, capped column
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...