Maintaining the COC Inlet
To Clean the COC Inlet
114
Maintaining Your GC
To Clean the COC Inlet
1
Gather the following:
•
1/4-inch and 5/16-inch wrenches
•
Cleansing bath
•
Aqueous detergent
•
Distilled water
•
Methanol
•
Compressed, filtered, dry air or nitrogen
•
Lint-free gloves
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3
Remove the column.
4
Remove the septum nut or cooling tower and then remove the septum.
5
Remove the existing insert from the inlet by gently pushing it out from below with a wire or
piece of column. Store the insert for possible later use.
6
Fill an ultrasonic cleansing bath with aqueous detergent and place the spring and insert
into it. Sonicate for 1 minute.
7
Drain the aqueous detergent and fill the bath with distilled water. Sonicate for 1 minute.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...