![Agilent Technologies 8890 Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678066.webp)
5
Maintaining the Split/Splitless Inlet
To Replace the Gold Seal on the Split/Splitless Inlet
66
Maintaining Your GC
6
Put on gloves to protect the new gold seal and washer from contamination. Put a new
washer in the reducing nut and place the new gold seal on top of it (raised portion facing
down).
7
Replace the reducing nut and tighten securely with a wrench.
8
Replace the inlet liner.
9
Install the column and the insulation cup.
10
“To Bakeout Contaminants from the Split/Splitless Inlet”
11
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.
12
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
13
Restore the analytical method.
Inlet gold seal
Washer
Reducing nut
Gold seal side view
Make sure raised portion faces
down.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...