Maintaining the PTV Inlet
To Clean the Septum Seat in the Septum Head Assembly of the PTV Inlet
Maintaining Your GC
155
To Clean the Septum Seat in the Septum Head
Assembly of the PTV Inlet
1
Gather the following:
•
Replacement septum, see
“Consumables and Parts for the Programmable
Temperature Vaporization Inlet”
•
5/8-inch wrench
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
•
Lint-free gloves
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
3
Remove the septum retainer nut or Merlin cap. If the septum head begins to turn during
removal, support it manually while removing the cap.
4
Unscrew the septum head assembly from the inlet and move it up and away from the inlet.
5
Use tweezers to remove the septum or Merlin Microseal from the retainer nut. Do not
gouge or scratch the interior of the septum head.
6
Scrub the residue from the retainer nut and septum holder with a small piece of rolled-up
steel wool and tweezers. Do not do this over the inlet.
7
Use compressed air or nitrogen to blow away the pieces of steel wool and septum.
8
Replace the septum head assembly on the inlet. Finger-tighten the septum head and then
tighten an additional 1/2 turn with a wrench.
9
Firmly press the new septum or Merlin Microseal into the fitting. See
10
Replace the septum retainer nut or Merlin cap and finger-tighten. See
11
Perform an inlet
Leak & Restriction
test and reset the maintenance counters.
12
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
13
Restore the analytical method.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...