7
Maintaining the COC Inlet
To Replace the Fused Silica Needle in a Syringe for the COC Inlet
118
Maintaining Your GC
To Replace the Fused Silica Needle in a Syringe for
the COC Inlet
NOTE
The fused silica needle and syringe are only used with the cooling tower and duckbill septum
for manual, on-column injections onto 200-µm columns.
1
Gather the following:
•
New fused silica syringe needle, see
“Consumables and Parts for the Cool On-Column
•
Solvent
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
2
Loosen the retaining nut and remove the old needle.
3
Hold the syringe vertically and insert the new fused silica needle so it is visible inside the
syringe barrel. If the needle cannot be inserted into the syringe barrel, the PTFE ferrule may
be blocked. You may need to replace the ferrule.
4
Push the plunger down until it bottoms. The needle will be flush with the plunger end.
5
Finger-tighten the retaining nut. Pull the needle gently to be sure the PTFE ferrule has
formed a tight seal with the needle. Tighten the retaining nut further, if necessary.
6
Loosen the retaining nut just enough so the needle is again free.
7
Depress the syringe plunger slowly until it pushes the needle to the end of the barrel, then
finger-tighten the retaining nut.
8
Use a solvent to rinse the syringe and check for leaks or blocks. Leaks may be fixed by
further tightening the retaining nut. Blocks or serious leaks require repeating this
procedure.
Fused silica needle
PTFE ferrule
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...