![Agilent Technologies 8890 Скачать руководство пользователя страница 187](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678187.webp)
11 Maintaining the FID
To Replace an FID Jet
Maintaining Your GC
187
To Replace an FID Jet
1
Gather the following:
•
Replacement jet
•
T-20 Torx screwdriver
•
1/4-inch nut driver
•
Tweezers
•
Compressed, filtered, dry air or nitrogen
•
Solvent that will clean the type of deposits in your detector
•
Clean cloth
•
Cotton swab
•
Lint-free gloves
2
Launch the GC maintenance wizard:
Maintenance > Detectors,
select the detector, then
select
Perform Maintenance > Replace FID Jet > Start Maintenance.
The wizard will walk
through the steps needed to replace the jet (those same steps are repeated below).
WARNING
Be careful! The oven and/or detector may be hot enough to cause burns. If the detector is
hot, wear gloves to protect your hands.
WARNING
Wear safety glasses to protect your eyes from flying particles while handling, cutting, or
installing glass or fused silica capillary columns. Use care in handling these columns to
prevent puncture wounds.
3
If installed, remove the capillary column from the detector.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
Remove the FID castle assembly and place it on a clean cloth.
5
Locate the jet inside the housing.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...