Maintaining the Purged Packed Inlet
To Install a Packed Glass Column
98
Maintaining Your GC
8
Withdraw the column 1 to 2 mm from both the inlet and detector. Finger-tighten both
column nuts.
CAUTION
Overtightening the column nut or forcing it to bottom in either the inlet or detector may
shatter the column.
9
Tighten both column nuts 1/4 turn with a wrench.
10
Press
Settings > Configuration > Columns
, then select
Columns 1
through
6.
Select
Column Type > Packed
, then choose your
Inlet
and
Outlet Connection
to identify the inlet
and detector to which the column is attached.
11 Scroll
down to manually set your column temperatures.
12
Exit maintenance mode. Select
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Finished.
13
Establish a flow of carrier gas and purge as recommended by the packing manufacturer.
Generally:
•
20 to 30 mL/min for 2-mm id glass or 1/8-inch od metal columns
•
50 to 60 mL/min for 4-mm id glass or 1/4-inch od metal columns
14
Restore the analytical method.
•
For FPD+, immediately turn off the flame.
•
For NPD, immediately turn off the bead.
15
After the GC becomes ready, wait 10 minutes then ignite the detector flame or adjust
offset on the NPD bead.
WARNING
Be careful! The oven, inlet, and/or detector may be hot enough to cause burns. If the oven,
inlet, or detector is hot, wear heat-resistant gloves to protect your hands.
16
Allow the oven, inlet, and detector to equilibrate at operating temperature, then retighten
the fittings.
17
Perform the inlet
Leak & Restriction
test and reset the maintenance counters.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...