Maintaining the PTV Inlet
To Clean the Septumless Head on the PTV Inlet
150
Maintaining Your GC
To Clean the Septumless Head on the PTV Inlet
1
Gather the following:
•
Syringe with 23-gauge needle, see
“Consumables and Parts for the Programmable
Temperature Vaporization Inlet”
•
Septumless head rebuild kit
•
Hexane
•
Clean, lint-free gloves
•
5/16-inch wrench
•
Lint-free gloves
2
Manually set the inlet and oven temperature to < 40
°
C, and wait for the inlet, oven, and
other parts you might come into contact with inside the oven, to cool before continuing.
Alternately, place the GC in maintenance mode:
Maintenance > Instrument > Perform
Maintenance > Maintenance Mode > Start Maintenance.
3
Disconnect the carrier gas line.
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If either is hot, wear
heat-resistant gloves to protect your hands
.
4
Unscrew and remove the septumless head assembly from the inlet.
5
Unscrew the sealing element from the head assembly and carefully remove the Viton seal
and the pressure spring
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...