![Agilent Technologies 8890 Скачать руководство пользователя страница 87](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678087.webp)
6
Maintaining the Purged Packed Inlet
To Change the O-Ring on the Purged Packed Inlet
Maintaining Your GC
87
To Change the O-Ring on the Purged Packed Inlet
1
Gather the following:
•
Replacement O-ring, see
“Consumables and Parts for the Purged Packed Inlet”
•
Septum nut wrench
•
Tweezers
•
Lint-free gloves
2
Launch the GC maintenance wizard:
Maintenance > Inlets,
select the inlet, then select
Perform Maintenance > Replace Liner and O-Ring > Start Maintenance.
The wizard will
walk through the steps needed to replace the liner and O-ring (those same steps are
repeated below).
WARNING
Be careful! The oven and/or inlet may be hot enough to cause burns. If the inlet is hot, wear
heat-resistant gloves to protect your hands
.
3
Loosen the top insert assembly to remove the top portion of the inlet.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
4
Use tweezers to remove the old O-ring.
5
Insert a new O-ring.
6
Install and tighten the top insert assembly.
7
The GC maintenance wizard will perform checks at the appropriate times, including
Leak &
Restriction
tests, and will automatically reset the maintenance counters.
8
Select
Finished
,
then select
OK
to exit the GC maintenance wizard
.
9
Restore the analytical method.
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...