![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552066.webp)
66
Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Install the Column Shelf
Install the Column Shelf
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Open the doors of the sampler
•
Open the ICC
WA R N I N G
Fire or explosion due to missing leak drainage
Solvents may run into the module if Column Shelf or Integrated Column
Compartment are not installed.
➔
Install either Column Shelf or Integrated Column Compartment before starting the
module.
1
Remove the capillary connection from the integrated
column compartment to the module.
2
Remove the column.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...