Agilent InfinityLab LC Series Vialsamplers User Manual
8
Error Information
Sample Cooler Error Messages
Cooler Condensate Error
Error ID: 30702, 30709, 30710
The sample cooler has detected condensate inside of the leak pane.
Probable cause
Suggested actions
1
Blocked drainage
Switch off the instrument and remove the
condensate tubings. Place a beaker underneath
the waste outlet. Check the waste container.
There should be no kinked and sealed tubing.
2
Defective condensate sensor
Switch the sample cooler off/on and check if
after 8 min the error shows up again.
3
Defective leak sensor
Replace the sample cooler.