Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Metering Seal and Piston
8
Use a seal insert tool to carefully remove the seal.
9
Clean the chamber with lint-free cloth and an
appropriate solvent. Ensure all particulate matter is
removed.
10
Use the plastic side of the insert tool to install the new
seal. Press the seal firmly into position.
11
Place the piston guide on top of the seal.
12
Reassemble the metering head assembly. Carefully
insert the piston into the base. The closed side of the
metering head must be on the same side as the lower
one of the two capillary drillings.
13
Install the fixing bolts. Tighten the bolts securely.
Tighten the bolts alternately two turns at a time until the
stator head is secure. The housing must be flush with
the metering head.
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...