Agilent InfinityLab LC Series Vialsamplers User Manual
275
Hardware Information
12
Interfaces
IO (Input/Output) Lines
•
Eight generic bi-directional channels (input or output).
•
Same as the APG Remote.
•
Devices like valves, relays, ADCs, DACs, controllers can be
supported/controlled.
1-Wire Data (Future Use)
This serial line can be used to read out an EPROM or write into an EPROM of a
connected ERI-device. The firmware can detect the connected type of device
automatically and update information in the device (if required).
Pin
Enhanced Remote
1
IO 1 (START
REQUEST)
2
IO 2 (STOP)
3
IO 3 (READY)
4
IO 4 (POWER ON)
5
IO 5 (NOT USED)
6
IO 6 (SHUT DOWN)
7
IO 7 (START)
8
IO 8 (PREPARE)
9
1 wire DATA
10
DGND
11
+5 V ERI out
12
PGND
13
PGND
14
+24 V ERI out
15
+24 V ERI out
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...