![Agilent Technologies 1260 Infinity II Series Скачать руководство пользователя страница 48](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552048.webp)
48
Agilent InfinityLab LC Series Vialsamplers User Manual
2
Site Requirements and Specifications
Specifications of the Sample Cooler
C A U T I O N
General hazards and improper disposal
Improper disposal of the media and components used pollutes the environment.
➔
The breakdown of the sample cooler or sample thermostat unit must be carried out
by specialist refrigeration company.
➔
All media must be disposed of in accordance with national and local regulations.
➔
Please contact your local Agilent Service Center in regard to safe environmental
disposal of the appliance or check
www.agilent.com
for more info.
Table 6
Performance Specifications Agilent 1290 Sample Cooler
Type
Specifications
Operating principle
High performance, low-energy consumption
micro-compressor based cooler with
ozone-friendly HFC-134a coolant (42 g),
user-upgradable.
Temperature range
from 4 °C to 5 °C below ambient
Temperature settable
from 4 – 40 °C in 1 ° increments
Temperature accuracy (<25 °C, <50 % r.H.)
2 °C to 6 °C at a setpoint of 4 °C
Содержание 1260 Infinity II Series
Страница 1: ...Agilent Technologies Agilent InfinityLab LC Series Vialsamplers User Manual ...
Страница 34: ...34 Agilent InfinityLab LC Series Vialsamplers User Manual 1 Introduction Leak and Waste Handling ...
Страница 56: ...56 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Turn on off 5 ...
Страница 88: ...88 Agilent InfinityLab LC Series Vialsamplers User Manual 3 Using the Module Agilent Local Control Modules ...
Страница 130: ...130 Agilent InfinityLab LC Series Vialsamplers User Manual 5 Preparing the Module Setting up the Vialsampler ...
Страница 306: ...306 Agilent InfinityLab LC Series Vialsamplers User Manual 13 LAN Configuration PC and Agilent ChemStation Setup ...