Agilent InfinityLab LC Series Vialsamplers User Manual
9
Maintenance and Repair
Exchange the Wash Port Assembly
8
Disconnect the sample loop from the needle.
9
Loosen the fixing screw.
10
Lift out the needle.
11
Disconnect the wash tubing from the peristaltic pump.
12
Release the spring loaded pin.
13
Lift the flapper out of the bracket. Be careful not to
squeeze the wash tubing.